The NHX Series is a UL 1069 Listed nurse call intercom. The system provides staff alert and communications for a variety of health care applications: nursing homes, assisted-living facilities, clinics, wards, and hospitals.
Aiphone is proud to be a rare, two-time winner of the prestigious Deming Prize for quality management and manufacturing processes. Our ISO Certified process and rigorous standards result in some of the most reliable products in the industry. Aiphone products routinely last for years, only being replaced when the application outgrows the system capacity.
Power Source
24V DC – use two PS-2420ULs
Communication
Handset (VOX) or push-to-talk (simplex)
Wire Type
Sub/Duty Station to Central Control Unit: 4 twisted pair, low cap, PE, overall shield – use Aiphone #862208
Master Station to Central Control Unit: 5 twisted pair, low cap, PE, overall shield – use Aiphone #862210
Distance
Sub Station to Central Control Unit: 165′(22AWG)
Duty Station to Central Control Unit: 65′(22AWG)
Master Station to Central Control Unit: 65′(22AWG)